Welch Allyn Medical Diagnostic Equipment Thermometer 692 Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Mesure, test et contrôle Welch Allyn Medical Diagnostic Equipment Thermometer 692. Welch Allyn Medical Diagnostic Equipment Thermometer 692 User's Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - SureTemp

Service ManualSureTemp® PlusModel 690 and 692 ThermometersPart No. 71221-0000 Rev. H

Page 2 - AGENCY APPROVALS

6 71221-0000 Rev. HFour probe types are available and interchangeable on the thermometer:1. A blue Oral/Axillary probe with a short (standard) coiled

Page 3 - Table of Contents

SureTemp® Plus Service Manual 72.2 Model FeaturesFeatures of the SureTemp® Plus Thermometer models (690 and 692) are listed below.2.3 Operating Featur

Page 4

8 71221-0000 Rev. HImmediately following the LCD segment test, the software revision is displayed on the screen for 2 seconds. At the end of the self

Page 5 - Emissions Summary

SureTemp® Plus Service Manual 92.3.5 Monitor ModeThe Monitor mode provides continuous temperature monitoring capability. The recommended wait time for

Page 6 - 1. General Information

10 71221-0000 Rev. H2.3.6 Pulse Timer Mode (Model 692 only)Pulse Timer mode is activated from Low Power or Recall mode by pressing the Pulse Timer but

Page 7

SureTemp® Plus Service Manual 112.3.9 F/C ConversionThe temperature switches to F or C scale by pressing and releasing the C/F button. The default t

Page 8 - 1.5 Classifications

12 71221-0000 Rev. H The following features and selections are available in the Biotech mode.Table 2- 3: Biotech Mode FeaturesFunction SettingDEVICE

Page 9 - 2. Operating Characteristics

SureTemp® Plus Service Manual 132.4 Instrument SecuritySecurity options include two different electrical security options that alarms after a user des

Page 10 - 2.1.6 Wall Mount

14 71221-0000 Rev. H3. Preventive MaintenanceThe following preventive maintenance is recommended to maximize uninterrupted service with the SureTemp®

Page 11 - 2.3 Operating Features

SureTemp® Plus Service Manual 153.4.2 Cleaning The Removable Probe Well• Remove the probe well from the unit. Unplug the latching probe connector to p

Page 12 - 2.3.3 Low Power Mode

© 2012 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is

Page 13 - 2.3.5 Monitor Mode

16 71221-0000 Rev. H3.6 Calibration TestingWhenever the clinical accuracy of the thermometer is in question, verify the instrument calibration. Your h

Page 14 - 2.3.8 Horn

SureTemp® Plus Service Manual 1711. Lift the rubber flap on the Model 9600 Calibration Tester and insert the probe fully into the small hole on top of

Page 15 - 2.3.11 Biotech Mode

18 Sure Temp® Plus Service Manual4. TroubleshootingVerify your Biotech Mode settings (page 11) to verify your device is configured properly. If the d

Page 16 - Function Setting

71221-0000 Rev. H 194.1.3 Patient Temperature Out of RangePatient Temperature Out of Range errors occur when the patient temperature is above 110.0° F

Page 17 - 2.4 Instrument Security

20 Sure Temp® Plus Service Manual4.1.6 Error Codes: TroubleshootingThe following table describes the error codes displayed on the thermometer, their c

Page 18 - 3. Preventive Maintenance

71221-0000 Rev. H 21Note: If your thermometer is within the warranty period, you should return the unit to an authorized service representative for s

Page 19 - 3.5 Replacing the Batteries

22 Sure Temp® Plus Service Manual4.2 Field Serviceable RepairsThe only operations that are performed by field personnel are calibration verification,

Page 20 - 3.6 Calibration Testing

71221-0000 Rev. H 235. Theory of Operation5.1 Technical OverviewThe SureTemp® Plus Model 690 and 692 are portable, battery powered thermistor-based m

Page 21

24 Sure Temp® Plus Service Manual5.1.1 Temperature ProbesModel 690 and 692 Thermometer probes are configured at the factory to be Oral/Axillary or Rec

Page 22 - 4. Troubleshooting

71221-0000 Rev. H 25Appendix A SpecificationsAmbient Operating Temperature Range 50.0 F to 104.0 F (10.0 C to 40.0C)Operating Altitude -557 f

Page 23 - 4.1.5 Instrument Errors

SureTemp® Plus Service Manual iTable of ContentsElectromagnetic Immunity and Emissions Declaration. . . . . . . . . . . . . . . . . . . . . . . . . .

Page 24

26 Sure Temp® Plus Service ManualAppendix B Limited WarrantyTHREE-YEAR LIMITED WARRANTY ON MODEL 692 AND TWO-YEAR LIMITED WARRANTY ON MODEL 690 I

Page 26 - 4.3 Field Serviceable Parts

71221-0000 Rev H4341 State Street RoadSkaneateles Falls, NY 13153Tel: +1 800 535 6663Fax: +1 315 685 3361www.welchallyn.com

Page 27 - 5. Theory of Operation

ii 71221-0000 Rev. H3.6 Calibration Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 28

SureTemp® Plus Service Manual 1Electromagnetic Immunity and Emissions DeclarationImmunity StandardEmissions SummaryThe SureTemp® Plus model 690 and 69

Page 29 - 71221-0000 Rev. H 25

2 71221-0000 Rev. H1. General Information 1.1 About this ManualThis manual describes maintenance and troubleshooting procedures for the SureTemp® Plu

Page 30

SureTemp® Plus Service Manual 3• Use single-use Welch Allyn disposable probe covers to limit patient cross-contamination. The use of any other probe c

Page 31

4 71221-0000 Rev. H• The Welch Allyn SureTemp Plus thermometer consists of high quality precision parts. Protect it from severe impact and shock. A qu

Page 32

SureTemp® Plus Service Manual 52. Operating Characteristics2.1 Basic System DescriptionThe Welch Allyn® SureTemp® Plus Model 690 and 692 thermometer